Safety chambers with vertical airflow

Laminar chamber K 500 series

Adjustable, horizontal laminar air flow of 0.20 to 0.50 m/s or according to guidelines.
Two-stage filtration system: polyamide pre-filter of EU3 (G3) class in accordance with EN 779 and HEPA filter H14.

Automatic filter wear compensation to ensure a constant and safe air flow rate in the chamber.
• Quiet fans, CA level according to PN-EN ISO 11201 below 49 dB.
Air velocity control by sensor or volt voltage system with ±10% user correction.
Two filters: pre-filter and main HEPA H14 with efficiency above 99.995% for MPPS.
UV germicidal lamp permanently mounted, placed in the upper, rear part of the working area, protected against accidental switching on during operation.
Hour counterof the working time of the device and the UV lamp of the chamber.
Energy-saving LED lampilluminating the workspace.
Intuitive touch panel control with menus in Polish, English, German and Spanish.
• Stainless steel worktop.
• Electrical socket on the chamber casing and one inside the chamber.
• Optional air, vacuum and gas valves and lockable castor base.

Technical details

External dim. (W x H x D) 515 x 500 x 815 mm
Internal dim. (W x H x D) 480 x 430 x 520 mm
Number of fans 1
Adjustable airflow 0,20 - 0,50 m/s
Number and type of filters 1 - main HEPA, 1 - pre-filter polyamide
HEPA filter efficiency >99,995 % dla MPPS - H14
Sound pressure PN-EN ISO 11201 < 55 dB
Lighting intensity > 600 luxów
Current consumption with LED fluorescent lamp 25 W
Power consumption when working without lighting 14 W
Number of electrical sockets opcjonalnie
Rated power ? W
Weight of the device with the base 36 kg

PRODUCT CERTIFICATES
• declaration of conformity CE
• the chamber is manufactured in accordance with the certified quality management system ISO 9001

Each chamber is tested in the company's research laboratory operating in accordance with the ISO 17025 standard.

During the test, the following parameters are checked:
• air flow velocity in the working space and barrier,
• number content of particles,
• homogeneity of the filter material (leakage test),
• tightness of the filter mounting,
• deflection of the laminar air stream.

A factory test report is issued and attached to each chamber.

Warranty of the laminar chamber: 36 months.